MIC 1422
MIC 1422MIC 1422
MIC 1422
MICROBASED CONTROLLER
MICROBASED CONTROLLERMICROBASED CONTROLLER
MICROBASED CONTROLLER
INSTRUCTION MANUAL
INSTRUCTION MANUALINSTRUCTION MANUAL
INSTRUCTION MANUAL
E-90
PN 136108
REVISION 10-07
Zobrazit stránku 0
1 2 3 4 5 6 ... 90 91

Shrnutí obsahu

Strany 1 - REVISION 10-07

i MIC 1422MIC 1422MIC 1422MIC 1422 MICROBASED CONTROLLERMICROBASED CONTROLLERMICROBASED CONTROLLERMICROBASED CONTROLLER INSTRUCTION MANUALINS

Strany 2

6 FIGURE 2-2 Main Dimensions MIC1422 FIGURE 2-3 Panel Mounting the Controller

Strany 3 - TABLE OF CONTENTS

7 2.2 Wiring Guidelines Electrical noise is a phenomenon typical of industrial environments. The following are guidelines that must be followed t

Strany 4

8 AC Power Wiring Neutral (For 115 VAC) It is good practice to assure that the AC neutral is at or near ground potential. To verify this, a voltme

Strany 5

9 Noise Suppression At The Source Usually when good wiring practices are followed no further noise protection is necessary. Sometimes in severe el

Strany 6

10 FIGURE 2-5 2.3 Sensor Placement (Thermocouple or RTD) Two-wire RTDs should be used only with lead lengths less than 10 feet. If the temperat

Strany 7

11 FIGURE 2-6 Wiring Label 1/4 DIN

Strany 8

12 2.4 Input Connections In general, all wiring connections are made to the instrument after it is installed. Avoid Electrical Shock. AC power w

Strany 9

13 FIGURE 2-7A 24V Nominal AC/DC Supply The supply connection for the 24V AC/DC option of the instrument are as shown below. Power should be conne

Strany 10

14 FIGURE 2-9 RTD Input Make RTD connections as illustrated below. For a three wire RTD, connect the resistive leg of RTD to terminal 1 and the co

Strany 11

15 FIGURE 2-11 Remote Digital Communications - RS485 Make digital communication connections as illustrated below. FIGURE 2-12 Remote Setpoint I

Strany 13

16 FIGURE 2-13 Remote Setpoint Selection Connections are made as shown. FIGURE2-14 Dual Setpoint Selection Connections are made as shown.

Strany 14

17 2.5 Output Connections FIGURE 2-15 Relay Output 1 (Control Output 1) Connections are made to Output 1 relay as illustrated below. The contact

Strany 15

18 FIGURE 2-18 Relay Output 2 (Control Output 2 OR Alarm 2) Connections are made to Output 2 relay as illustrated below. The contacts are rated at

Strany 16

19 FIGURE 2-21 Relay Output 3 (Alarm 1) Connections are made to Output 3 relay as illustrated below. The contacts are rated at 2 amp resistive, 12

Strany 18

21 SECTION 3: OPERATION 3.1 Control Capability The capabilities available in a specific unit are dependent upon the hardware options specified w

Strany 19

22 3.3 Direct/Reverse Operation of Outputs Direct operation is typically used with cooling applications. On-Off direct output(s) will turn on whe

Strany 20

23 When the unit is operating in the Control Mode, the control algorithm determines the output % required to correct for any difference between the

Strany 21

24 FIGURE 3-1 Proportional Band 1

Strany 22

25 3.7 Power Up Procedure Verify all electrical connections have been properly made before applying power to the instrument. If the instrument is

Strany 23

iii TABLE OF CONTENTS TABLE OF CONTENTS...iii SECTIO

Strany 24

26 UP KEY • Increase the displayed parameter value. • Increase setpoint. DOWN KEY • Decrease the displayed parameter value. • Decrease setpoi

Strany 25 - SECTION 3: OPERATION

27 SECTION 4: CONTROL MODE 4.1 Operation After the instrument has performed its power up self test, the Control Mode is active with the setpoint

Strany 26

28 4.3 Override Feature While the instrument is being used with either Dual Setpoint operation or Remote Setpoint operation, the Override feature

Strany 27

29 If a break is detected in the sensor circuit, the upper display will show: 4.6 Manual Control (Percent Output) Manual Control is not applica

Strany 28

30 4.7 Setup Modes The Setup Modes contain parameters which configure the instrument and affect how the control functions. To access the Setup Mod

Strany 29

31 SECTION 5: TUNE MODE 5.1 Tune Mode Description The Tune Mode contains parameters concerning tuning of the instrument. To access the Tune Mod

Strany 30

32 5.2 Manual Tuning Method 1. Cycle Time - Time Proportioning Outputs A. Adjusting the cycle time affects instrument operation 1. Shorter Cyc

Strany 31 - SECTION 4: CONTROL MODE

33 Table 5-1 Tune Mode Parameters STEP DESCRIPTION DISPLAY CODE AVAILABLE SETTINGS FACTORY SETTING 1 Local Setpoint LSP +/- Setpoint Limits I

Strany 32

34 FIGURE 5-1 Proportional Band & Deadband/Overlap

Strany 33

35 SECTION 6: ALARM MODE 6.1 Alarm Mode Description The Alarm Mode contains parameters concerning process alarms. To access the Alarm Mode from

Strany 34

iv SECTION 7: TEST MODE ... 43 7.1 Test Mode Description .

Strany 35 - SECTION 5: TUNE MODE

36 Table 6-1 Alarm Mode Parameters STEP DESCRIPTION DISPLAY CODE AVAILABLE SETTINGS FACTORY SETTING 1 Alarm 1 Type ALA1 P-hi=Proc High nonE

Strany 36

37 FIGURE 6-1 Alarm Actuation

Strany 38

39 FIGURE 6-2 Alarm Hysteresis

Strany 39 - SECTION 6: ALARM MODE

40 6.2 Loop Alarm Enable This parameter is the means by which the user can enable or disable the Loop Alarm. The Loop Alarm is a special alarm wh

Strany 40

41 6.4 Logical Combination of Alarms Two alarms may be combined logically to create an AND/OR situation. They may be configured for Reverse-acting

Strany 41

42 FIGURE 6-3 Asymmetrical Band Alarm

Strany 42

43 SECTION 7: TEST MODE 7.1 Test Mode Description The Test Mode allows manual control of the instrument outputs in order to test their operation

Strany 44

45 SECTION 8: CONFIGURATION MODE 8.1 Configuration Mode Description The Configuration Mode contains parameters concerning output functions of th

Strany 45

1 SECTION 1: PRODUCT DESCRIPTION 1.1 General This instrument is a microprocessor based single loop controller capable of measuring, displaying an

Strany 46

46 STEP DESCRIPTION DISPLAY CODE AVAILABLE SETTINGS FACTORY SETTING 5 Output 2 Usage USE2 Out2=Control (opposite of Out1 action) Hy_r=Alm

Strany 47 - SECTION 7: TEST MODE

47 * The Hardware Definition Code and input jumper configuration may need to be changed. See Appendices A and B. ** If Remote Setpoint Input has

Strany 48

48 8.2 Hardware Definition Code The Hardware Definition Code is used to represent the hardware installed (input type, Output 1 type, Output 2 type

Strany 49

49 The displayed code may be incremented/decremented using the UP/ DOWN keys as required. The maximum setting available is 4777. For example, the

Strany 50

50 NOTE: It is essential that this code is changed whenever there is a change to the instrument's hardware configuration (change of input/outp

Strany 51

51 SECTION 9: CALIBRATION MODE 9.1 Entering Calibration Mode To enter the Calibration Mode from Setup Mode: 1. Press the SCROLL key until CAL a

Strany 52

52 3. Enter Calibration Mode (Section 9.1). The upper display will then show Input Type Number, in the form: iP_I and the lower display will show:

Strany 53

53 9.3 Calibrating The Secondary Analog Input Equipment Required 1. DC linear input source (0 - 5 V and 0 - 20 mA) with an accuracy better than

Strany 54

54 6. To calibrate all inputs, repeat Steps 1 to 4 for each of the other input types (see Table 9-2) until all three secondary analog input types h

Strany 55 - SECTION 9: CALIBRATION MODE

55 SECTION 10: ENABLE MODE 10.1 Enable Mode Description The Enable Mode controls access to the various modes available. To access the Enable Mo

Strany 56

2 1.2 Displays Each instrument is provided with dual displays and status indicators as shown in Figure 1 -1. The upper display (RED) displays th

Strany 57

56 Table 10-1 Enable Mode Parameters STEP DESCRIPTION DISPLAY CODE AVAILABLE SETTINGS FACTORY SETTING 1 Enable Setpoint Change ESP EnAb = Enabl

Strany 58

57 SECTION 11: PRE-TUNE MODE 11.1 Pre-Tune Mode Description The Pre-Tune Mode may be used to set the instrument's PID parameters to values

Strany 60

59 SECTION 12: AUTOTUNE MODE 12.1 AutoTune Mode and Description The Auto-Tune Mode is used to optimize tuning while the instrument is operating.

Strany 61 - SECTION 11: PRE-TUNE MODE

60 New instruments supplied by the factory contain PID terms set at "DEFAULT" values which have been found to give adequate and safe contr

Strany 62

61 SECTION 13: RaPID FEATURE 13.1 RaPID Description The RaPID (Response assisted PID) range of controllers have been designed with a unique &quo

Strany 64

63 APPENDIX A: BOARD LAYOUT - JUMPER POSITIONING FIGURE A-1 OUTPUT 2, OUTPUT 3 REMOVAL

Strany 66

65 FIGURE A-3 PSU PWA WITH RELAY OR SSR OUTPUT 1

Strany 67

3 FIGURE 1-1 Keys and Indicators 1.5 Process Variable/Setpoint Value Re-Transmission Output If the instrument is specified with this option, thi

Strany 68 - FIGURE A-2 CPU PWA

66 FIGURE A-4 PSU PWA WITH DC OUTPUT 1

Strany 69

67 FIGURE A-5 OPTION PWA DC OUTPUT 2/OUTPUT 3

Strany 70

68 FIGURE A-6 JUMPER PLACEMENT FOR REMOTE INPUT TYPE

Strany 71

69 APPENDIX B: RANGE CODES The input ranges available (selectable via the front panel) are: For Thermocouple Inputs TYPE INPUT RANGE DISPLAYED CO

Strany 72

70 For DC Inputs Note: Input conditioning jumper LJ1, LJ2. or LJ3 needs to be changed, see Appendix A. INPUT RANGE DISPLAYED CODE 0-20mA 3413 4-2

Strany 73 - APPENDIX B: RANGE CODES

71 APPENDIX C: RaPID CONTROL FEATURE The RaPID (Response - assisted PID) feature offers dramatic improvements in control quality compared with con

Strany 74

72 In conditions of frequent change in load characteristics, it is recommended that the Auto-Tune facility is used. Note: With Auto-Tune and RaPID

Strany 75

73 APPENDIX D: SPECIFICATIONS Input Specifications General Input Sample Rate: Four per second Input Resolution: 14 bits approximately In

Strany 76

74 Dual Setpoint Selection Input Type: Voltage free or TTL compatible Voltage Free Operations: Connections to contacts of external switch

Strany 77 - APPENDIX D: SPECIFICATIONS

75 To Select Remote Setpoint/ Setpoint 1: Minimum contact resistance (open): 5K ohms Minimum voltage for (TTL) for " 1 ": 2.0 V Maximum

Strany 79

76 Output 2 General Types Available: Relay, SSR and DC Relay Contact Type: Single pole double throw (SPDT) Rating: 2A resistive at 120/240V A

Strany 80

77 DC Resolution: Eight bits in 250 mS (10 bits in 1 second typical, >10 bits in >1 second typical). Update Rate: Four times per second R

Strany 81

78 Alarms Maximum Number: Two "soft" alarms plus Loop Alarm* Maximum # Outputs: Up to 2 outputs can be used for alarm purposes Combina

Strany 82

79 DC Outputs Output 1 Accuracy: mA: 0 - 20mA ± 0.5% of span (20 mA) @250 ohm 4 - 20mA ± 0.5% of span (16 mA) @ 250 ohm V: 0

Strany 83

80 Performance Under Operating Conditions Temperature Stability: 0.01 % of span /degree C change in ambient temperature Cold Junction Compensation

Strany 84

81 APPENDIX E: SOFTWARE REFERENCE SHEET Tune Parameter Setting LSP rSP SP1 SP2 iCor Po1 Po2 Pb1 Pb2 ArSt rAtE rSEt HyS1

Strany 85

82 Alarm Parameter Setting ALA1 ALA2 Inhi (Alm 1 Value) AHy1 (Alm 2 Value) AHy2 LAEn LAti Enable Parameter Setting ESP EPrE

Strany 86

83 Configuration Parameter Setting InPS FiLt rinP CtL1 USE2 USE3 CbS CPAr CAd CJC dPoS Euu EuL SPuL SPLL rSPu rSPL

Strany 87

84 APPENDIX F: FLOW CHART OF OPERATION

Strany 88

85 APPENDIX G: MODEL NUMBER MATRIX

Strany 89

5 SECTION 2: INSTALLATION AND WIRING 2.1 Mounting Electrical code requirements and safety standards should be observed and installation performe