WORKING WITH OPERATING MODES Protocol 3 Controller Owner’s Manual
32 Version 1.2
Copyright © 2013 by Despatch Industries.
All rights reserved. No part of the contents of this manual may be reproduced, copied or transmitted in any form or by any
means including graphic, electronic, or mechanical methods or photocopying, recording, or information storage and
retrieval systems without the written permission of Despatch Industries, unless for purchaser's personal use.
Table 12. Profile Setup Steps, Options and Descriptions.
Disallow a remote source to run a profile
Allow a remote source to run a profile
Disallow use of delayed start time
Allow controller to start at a later time with
programmed delay
Unique name for each profile using up to 8
characters
Profile start not delayed
Set hours and minutes to delay before
starting the profile
Set time of day (hour and minutes) and day
of week to start profile. Options include
each day of the week and combinations of
days.
If power-down occurs while this profile is
running, specify the action for the controller
to take during power-up:
Controller restarts current profile
Controller maintains previous profile
setpoint
Controller continues (for specified time) the
profile from where it was when power
failed.
If program aborts, profile should:
Choose the number of times to cycle the
program, up to an infinite number of times.
Set target setpoint (degrees centigrade)
Set the segment ramp time (hh:mm:ss)
Set the Auto-Hold Type: None, Above
Setpoint, Below Setpoint, Band.
Set Events 1-5: Press or to
toggle between Active and Inactive.
Set the target setpoint (degrees
centigrade)
Set the segment ramp rate in units per
hour
Set the Auto-Hold Type: None, Above
Setpoint, Below Setpoint, Band.
Set Events 1-5: Press or to
toggle between Active and Inactive.
Set the target setpoint (degrees centigrade)
Choose time to dwell at setpoint.
Specify Auto-Hold type: None, Above
setpoint, Below setpoint, Band