If active, Pre‐tune will attempt to run at every power up.
Automatic tuning will not engage if proportional band is set to
ON/OFF control.
Pre
‐
tune and Auto Pre
‐
Tune will not engage if the setpoint is
ramping or the Process Variable is < 5% of span from setpoint.
5.8. Product Information
Access Product Information through the Main Menu (Section 5.2.4). Product Information
provides detailed information on the unit’s software and hardware, such as calibration, firmware
and production revision. Table 18 detailed description of the information is provided in the table
below.
Press to step through and accept all product information options.
Table 18. Product Information Details.
Calibration status of mVDC, VDC, mADC, RTD and
Thermocouple CJC inputs. All should display as Calibrated
USB Port; Data Recorder (includes USB Port) or Profiler.
Type and version of firmware.
Software and Hardware revision levels
Serial Number Information
Instrument serial number.
5.9. Service Information
Access Service Information (Figure 20) through the Main Menu (Section 5.2.4). The Service
Information menu displays contact information for service, sales or technical support.